The following are several criteria necessary to consider
when selecting an ATR crystal material for a specific
sample:
Refractive Index - the crystal should have a higher index of refraction
than the sample. The majority of organic samples have refractive indices
in the area of 1.5. Refractive indices of standard ATR crystals span
from 2.4 to 4.0 - which in most cases provides sufficient sample to crystal
differentiation. Inappropriate refractive index ratios may cause distortion
of spectral features. These may be manifested by diminished peak symmetries,
sharp baseline/peak shoulder transitions, and in extreme cases, presence
of derivative-like features in the spectrum.
Spectral Range - all ATR crystals have different spectral ranges. Specifically,
in mid-IR the cutoff at low wavenumbers varies from approximately 780
cm-1 for Ge to 250 cm-1 for KRS-5. To a certain extent, the cutoff values
are also affected by the length (thickness) of the crystal. In light
of these facts, it is important to determine whether the spectral features
of the sample correspond with the spectral range of the ATR crystal selected.
Useful spectral ranges of the most popular ATR materials are listed in
the table below.
Chemical and Physical Properties - for obvious reasons, the ATR crystal
must be chemically and physically compatible with the sample. Some crystal
materials may react with samples. This will typically damage the crystal
surface and may produce unpleasant side effects (e.g. acidic solutions,
pH<5, may etch the ZnSe crystal, strong acids will generate toxic
hydrogen selenide). Physical considerations are equally important as
some crystals are more susceptible to pressure and temperature changes
than others. Selected precautions are listed in the table below.
| |
n1 |
LWL, cm-1 |
dp |
Water Solubility, g/100g |
pH Range |
Hardness, Kg/mm2 |
| AMTIR |
2.5 |
625 |
1.46 |
Insoluble |
1-9 |
170 |
| Diamond/ZnSe |
2.4 |
525 |
1.66 |
Insoluble |
1-14 |
5,700 |
| Diamond/KRS-5 |
2.4 |
250 |
1.66 |
Insoluble |
1-14 |
5,700 |
| Germanium |
4 |
780 |
0.65 |
Insoluble |
1-14 |
550 |
| KRS-5 |
2.37 |
250 |
1.73 |
0.05 |
5-8 |
40 |
| Silicon |
3.4 |
1500 |
0.84 |
Insoluble |
1-12 |
1150 |
| Silicon/Znse |
3.4 |
525 |
0.84 |
Insoluble |
1-12 |
1150 |
| ZnS |
2.2 |
850 |
2.35 |
Insoluble |
5-9 |
240 |
| ZnSe |
2.4 |
525 |
1.66 |
Insoluble |
5-9 |
120 |
| n1 = refractive index of ATR crystal |
| LWL = long wave length cut-off |
| dp = depth of penetration in microns @ 1000
cm-1 |
Sensitivity - effective pathlength of the infrared beam in the
sample must be sufficient to produce an adequate spectrum. This parameter
is affected by the number of reflections (more reflections yield higher
absorbance), and the depth of penetration - which is a function the refractive
indices and the angle of incident beam. Most ATR accessories are already
optimized to provide the best possible sensitivity. Standard configurations
provide 9 to 20 reflections at 45 degree angle of incidence. Changing
these parameters, which is possible with variable angle ATRs (or special,
dedicated crystal plates) can improve results in some cases. However,
when selecting special configurations the following issues should be
considered:
- A higher angle of incidence results in less reflections, and decreased
penetration depth, lowering the overall absorbance of the spectrum.
This is useful when highly absorbing or high refractive index samples
are being measured.
- A lower angle of incidence results in more reflections and an increased
penetration depth. This results in the best sensitivity but can result
in band distortions due to the large variation in sample refractive
index in the region of absorbance bands (anomalous dispersion).
Optical Design - the overall optical design of an HATR accessory
- its optical path, mirrors quality and throughput has great effects
on analytical results. Placing a good quality HATR accessory with a 45
degree ZnSe crystal in the sample compartment of an FTIR spectrometer
should result in an energy throughput of 25 to 40 % T. To verify performance
of an ATR accessory, collect the background spectrum with an empty sample
compartment. Place the accessory in the spectrometer and collect the
transmission spectrum. The resulting spectrum should be in the 25 to
40% T range.
|